CAN I SEE NANO-ORDER?

There is a limit to what can be seen with visible light. Light has a wavelength (length), and those shorter than the wavelength of light can not be seen by the eye. There are limits of optical microscope / microscope there.

Therefore, an electron microscope using an electron beam (electron beam) whose wavelength is shorter than visible light.

 

In the past, there were only 10 million yen or more for digital micrometer microscope. Right now, some digital micrometer were about 6 million yen.

 

In various regions, there are public industrial testing laboratories, some of which offer testing services and equipment usage with electronic microscopes.

 

※Please note that we do not handle electron microscopes.

WHAT IS THE DIFFERENCE BETWEEN A SIMPLE XY TABLE AND DIGITAL XY TABLE?

It is a support table (sample table) that can be moved in both X and Y directions. It is advantageous when using the appropriate position adjustment becomes easier.

 

A simple XY table that can move the stage with a coarse movement knob.

A simple XY table that can move the stage with an auxiliary knob. Xy table can move in large space and wide range. In addition, XY table can move quicky since the amount of movement per dial rotation is high.

 

A digital micrometer XY table is a XY table attached to the microscope in both X and Y directions. XY stage can move precisely with fine movement by turning the knob of the digital micrometer. The length can also be measured from the photographed image and the numerical value of micrometer. It is particularly effective for measuring sizes that can not be measured with image measurement software. Also, since the digital micrometer is capable of issuing calibration certificates, it is suitable for high precision measurement of length.

Although it can move precisely, on the other hand, it moves only 0.5 mm in one turn. Therefore, you can not move quickly.

 

簡易XYテーブル(落射照明タイプ) TK100 Simple XY table TK100
デジタルマイクロメータ付きXYテーブル Digital micrometer XY table TD100
デジタルマイクロメータ付きXYテーブル Digital micrometer XY table TD100 can measure length with high accuracy

 

If you need any furthure information, please don’t hestitate to contact our Technical Advisor.

WHAT’S VIGNETTE?

What is vignette?

 

A black part like a black tunnel which occurs when using an incompatible lens with image ssensors. It looks like it peeps in the tunnel.

ケラレ

See the dark side in the 4 corners of the frame

 

Cause of VIGNETTE

For example, in the case of a borescope,
As the sensor size of the camera connected to the borescope becomes larger, the vignetting area becomes wider. On the other hand, the image seen directly with the naked eye becomes closer to the image observed through the borescope.

 

In the case of a microscope
Typically, microscope cameras are designed not to have vignette (shadows in the four corners). Designers make allowances to prevent vignette, but the degree varies depending on the manufacturer.

WHAT’S MEGAPIXEL?

Megapixel means 1.000.000 pixels.

Megapixel camera refers to cameras with more than 1 million pixels.

Besides that, there are also lens that corresponds to Megapixel.

If you choose the lens that matches with the number of pixels of the camera, the picture will be better.

The balance between camera and lens is very important.

If you have any requirements on camera and lens, please don’t hestitate to contact us.

WHAT’S SCREEN RESOLUTION?

Pixel density in a bitmap image refers to the numerical representation of the granularity of the grid that expresses the image, indicating how many divisions there are in one inch.

 

In the case of a microscope, each of the camera, lens, and monitor has its own resolution.

 

1. Camera and lens resolution

Camera and lens resolution refers to the granularity of the grid that expresses the image.

 

It is a term that represents the detailing capability for subjects (such as screens, print results, photographic films, etc.) where the physical size of the image can be defined.

 

Therefore, physical resolution expresses the ability to distinguish the smallest unit of that image (pixels on a screen, resolution lines/spaces on print results or photographic film) per unit length of the subject, and it represents the level of detail rather than just the total count.

 

When selecting a microscope, especially the resolution of the camera and lens is crucial. For instance, if the camera has high resolution but the lens has low resolution, the resulting image will have poor resolution. Therefore, please be mindful of this aspect when choosing a microscope.

 

2. Monitor resolution

 

Monitor resolution refers to the ‘total number of pixels used for display,’ such as 1024×768.

 

XGA: 1024×768 pixels

SXGA: 1280×1024 pixels

VGA: 640×480 pixels

 

Description

 

Resolution

 

Monitor ratio

VGA

640×480

4:3

SVGA

800×600

4:3

XGA

1024×768

4:3

WXGA

1280×768

15:9

WXGA

1280×800

16:10

SXGA

1280×1024

5:4

UXGA

1600×1200

4:3

FHD (Full-HD, 1080p)

1920×1080

16:9

 

If you are using our PC Monitor Direct Microscope, Video Microscope, and High-Definition Microscope, the monitor aspect ratio is crucial.

 

For the PC Monitor Direct Microscope TG80XGA3 and all Video Microscope models, please use a monitor with a 4:3 aspect ratio or a widescreen monitor that can switch to a 4:3 aspect ratio.

 

For the High-Definition Microscope, use a monitor with Full High-Definition support (1080p) and for the 4K Microscope, use a monitor with 4K support and a 16:9 aspect ratio, or the displayed images may not have the correct proportions.

 

For USB microscopes, you can use them regardless of the monitor’s resolution. The USB camera’s image sensor has a 4:3 aspect ratio, so even if the PC screen has a different aspect ratio, the image within the launched application will be in a 4:3 format.

How to choose a Digital microscope?

The selection criteria for microscopes can be broken down into three main steps:

 

 

STEP 1: Basic System Selection

1. Magnification (Field of View): Consider the range of magnification options based on the desired field of view.
– Low Magnification: 5–50x (Approximately 67.5 x 51.0mm to 7.0 x 5.3mm field of view)
– Medium Magnification: 20–140x (Approximately 17.5 x 13.2mm to 2.7 x 2.0mm field of view)
– High Magnification 1: 40–240x (Approximately 9.0 x 6.7mm to 1.4 x 1.0mm field of view)
– High Magnification 2: 70–800x (Approximately 4.5 x 3.2mm to 0.4 x 0.3mm field of view)
– Ultra-High Magnification: Beyond 1000x (Approximately 1.5 x 1.1mm to 0.12 x 0.09mm field of view)

2. Working Distance: Consider the distance from the lens tip to the object, which is fixed for zoom lenses but varies with magnification.

3. Display Speed: Choose between High-Definition (HD) for smooth real-time imaging and USB for various additional features.

 

 

STEP 2: Resolution Selection

– HD Type: Choose between Full HD (1920×1080) or 4K (3840×2160) resolution.
– USB Type: Select from cameras with resolutions such as 1.3 million pixels (CMOS) or 5 million pixels (CMOS), considering factors like real-time display and data transfer speed.

 

 

STEP 3: Lighting Selection

– Ring Lighting is the basic option, but for higher magnifications, choose between Ring Lighting and Coaxial Lighting.
– Additional options like Transmitted Lighting, Dome Lighting, and Low-Angle (Darkfield Observation) Lighting are available.

 

 

Additional Considerations

– Explore optional accessories for diagonal observation, polarized observation, height measurement, auxiliary lenses, rear converters, filters, and other optical options based on specific needs.

 

 

Conclusion

 

Microscope selection involves considering magnification, working distance, display speed, resolution, and lighting. Evaluate these factors to meet specific requirements and applications.

Types of microscopes used for observing oral bacteria

Phase contrast microscopy is used for observation of oral bacteria including periodontal disease bacteria.

Phase contrast microscope is a biological microscope that can observe by converting the phase difference of light into contrast.

You can observe specimens close to transparent bodies without dyeing.

 

By using a normal phase contrast microscope you can observe bacteria in the oral cavity.

 

There seems to be a company specialized especially for dental clinics especially among them.

http://www.ptech.jp/pscope/(web site in Japanese)

http://www.ipona.biz/dentist/dentist.html(web site in Japanese)

Confirmation points if you thought that the light intensity of metallic microscope is dark

Metallic microscopes have more adjustment points than ordinary microscopes. Depending on the adjustment, the light intensity may decline extremely. If it is dark (or no light comes out) please check as the following instructions.

 

Polarization observation change lever

When both levers protrude, the light path is blocked.

   

Be make sure to switch between polarization observation and normal observation.

   
■Optical axis adjustment of light source

If the optical axis and the tilt of the light source deviate greatly, the light irradiated from the objective lens may be drastically declined.

   
■Adjusting the aperture in the middle of the optical path  

There are two aperture functions in the middle of the light path.

If this is narrowed too much, the irradiated light may decay considerably.

 

How to connect microscope camera to C mount?

The microscope camera has a connector len at the top.

Therefore, you can attach to the eyepiece part of the microscope and the JIS lens.

connect to eye piece part connect to Trinocular lens barrel
(camera port)

 

You can also connect to the C mount by removing the connector lens at the tip.

Camera for the microscope consists of the three parts of the belows photo.

 

  1. Camera body 2. C mount ring 3. Connector lens

Remove only the connector lens, be make sure to attach the C mount

 

In this picture, you can connect as a C  mount camera.

 

In this case, the lens is not attached to the camera. Therefore, you need to adjust the field of view on the C mount side (microscope side).

For details, please refer (“Connection of camera for microscope”).

How to integrate with binocular stereomicroscope?

When observing with a binocular stereomicroscope with both eyes, the objective lens is slightly oblique so that the object is close to the actual view.

(It is the same as looking at nearby objects with both eyes of a human.)

 

Therefore, looking at the following objects with one eye

 

 

You can see the side slightly. (In other words, you will be looking at diagonal.)

 

When the microscope is moved up and down, the field of view also shifts slightly.

The focus synthesis software is made on the premise that the field of view does not shift.
(It is assumed that it vertically moves up and down to the object)
If you use focus synthesis software with a stereomicroscope, it will blur like a photo below.

 

Mitani Shoji’s focal synthesis software (WinROOF 2018 Lite) has “microscope mode”.

Focus synthesis is possible without bleeding by focus synthesis using this function.

We also support these problems. Please feel free to contact us.

 

Of course, if you are a straight tube microscope like the one on the left, you can use normal focus synthesis software without problems.

Matching the focal point of the naked eye observation and the focal point of the microscope camera

Normally, focus adjustment of the microscope is performed when observe by naked eye.

Due to individual variations of the camera, the focal point of the naked eye observation may not match the focus of the camera.

In the case of JIS lens (see “Connecting camera for microscope”), the back focus is attached to the lens.
You can adjust easily.

 

In case of C mount, some of the microscope side C mount part has back focus, but it is very slight.

There is also a mechanism to adjust the back flange (back focus in this case) on the camera side. However, I think that this is also numerically small.

 

What is an eyepiece micrometer (reticle)?

Eyepiece micrometer (reticle) is a glass scale used by incorporating it in an eyepiece when measuring the size of a specimen with a microscope.

There are various kinds depending on how to use it.

Corresponding size varies depending on the size of the microscope.

φ19, φ20, φ20.4, φ21, φ24, φ25, φ26, φ27 mm are general-purpose sizes and can be easily purchased with onlineshops.

For many eyepieces, you can insert later as needed.

レチクル

Since there are some errors in the magnification of the objective lens, the measurement of eyepiece micrometer (reticle) becomes simple.

<How to use>

The width of the observed objects on a bar divides will be changed depend on the magnification of the objective lens.

Caculation:

Width on specimen of 1 scale (mm) = actual size of 1 scale of micrometer / magnification of objective lens

<Example>

Using a micrometer (0.1 mm / scale) of 10 mm / 100,

When a lattice sample of 0.2 mm pitch was measured with a microscope has objective magnification of 10 times

0.1 mm (actual scale of 1 scale) / 10 times (objective magnification) = 0.01 mm (1 scale on sample)

 

 < Measurement 10mm/100 (reticle) has step 0.1mm>

 

レチクル レチクル
< Measurement 10mm/100 (reticle) has step 0.1mm>

 

Measure sample of 0.2 mm lattice

 

レチクル レチクル

 

In 0.2 mm has 20 scales. (Precisely 19.5 scales)

The difference (about half scale) is due to the magnification error of the objective lens.

CAMERA CONNECTION FOR MICROSCOPES

The camera port of the microscope is roughly divided into the following two types: JIS lens and C mount

(Besides that, there are other classic or foreign types)

JIS lens

(straight lens diameter of 23.2mm)

C-mount

(Screw M25.4mm)

   

If you want to connect a camera with a “C-mount” to the mount of a microscope, there are two methods:

Convert the camera port on the microscope side to C-mount:
If the microscope already has a C-mount, you can directly attach the camera to it.

However, if the microscope has a JIS (Japanese Industrial Standard) microscope tube, you will need to modify the port on the microscope side.

Please note that not all microscopes allow for port conversion, so it may not be possible in some cases.

 

 

 

 

The biggest advantage of this camera is to be connected to the eyepiece section.

Cマウント

 

 

 

 

<Camera for microscope attached to C mount>

C mount is one of mount camera standards. (Please refer to “C mount, CS mount”)

Because of mount camera standards, a microscope camera is unnecessary.

It is the greatest merit that C mount type general purpose camera can be connected as it is. (Can not be connected to eyepiece section.)

 

 

However, the field of view (magnification) will change depending on the size of the image sensor on the camera.

As there is no lens on the camera, the magnification can not be adjusted.

(It must be selected when purchasing the camera adapter for microscope)

The following picture is a system diagram of some microscope manufacturers.

C-mount has various types from 0.25X to 0.63X.

It is necessary to purchase C mount matches to the number of inches of C mount camera.

Since it’s all C mount, the focus themselves matches. However, the field of view will be narrow or vignetting (shadows of four corners of the image) occurs.

 

2.  Make a C-mount camera compatible with a JIS (Japanese Industrial Standard) microscope tube

JIS鏡筒

When you attach a JIS microscope tube adapter lens to your camera, you can mount it onto the JIS microscope tube.

The biggest advantage of this camera is that it can also be connected to the eyepiece lens section.

If you remove the front lens, you can also connect it to a C-mount. Please refer to the instructions on “How to Connect a Microscope Camera to a C-mount” for reference.

3. Field of View Considerations:

 

Camera internal sensors (CMOS or CCD) come in various sizes. Ideally, the lens magnification should be adjusted according to the sensor’s size.
If the sensor size is 1/2 inch, it is recommended to choose a lens with approximately 0.5x magnification.

 

 

Cマウント JIS鏡筒

If you have other inch sizes or specific requirements, please contact our technical support.

 

Field of view when taking Full HD photograph by microscope

High definition video will be horizontally image of 16: 9. (1920 X 1080)

 

The field of view of the stereo microscope is a round view.

 

To take pictures of a microscope with a Full HD camera, there are two following methods.

1.

If shooting all fields of view of the microscope, put out small images on the monitor.

(The resolution will be about ¼)

    2.

If shooting to avoid dark conrner,

It will be smaller than the actual field of view.

(In fact, it is enlarger)

it will be smaller than the field of view

 
     
Field of view observe with the naked eyes   HDCE-20T
Full HD camera images for the HDCE-20T microscope
 
     

 

If you want to extend the field of view, you should use a 4: 3 camera (PC monitor direct with (analog RGB) camera or video camera) and se a len with a wide field of vision. (Some vignette will come out.)

 

Below picture is microscope camera with X 0.36 lens attached to our PC monitor direct (analog RGB) camera.

* Shadows appear in the four corners. Resolution is also lower than Full HD.

 

Use a 4: 3 camera and a wide field of vision len, the image likes the picture below.

How to save observation images with a microscope?

I will explain the field of view of a camera of microscope attached to an eyepiece lens or JIS lens.

(For camera of microscope which C mount is attached, please refer to “Connection of microscope and C mount camera”.

 

The left picture shows the image using microscope with the naked eye.

1. Camera of microscope  

Normally, cameras for microscope are made in order to avoid vignettes (shadows of the four corners)

It is the red frame of the top.

Although it designs with margin in order to avoid vignetting, its level of variation depending on the manufacturer.

However, it is enlarged more reliably than the naked eye observation.

Because the magnification of the lens of camera for microscope is fixed, it is not possible to change the field of view.

   
2. Commercial video camera and connector lens

I am observing it by installing a connector len on a commercially video camera.

Commercially video cameras have a zoom function.

Using this zoom function, we adjusted the field of view.

Vignetting (shadow of four corners) occurs, but you can take a wide field of view. (Yellow frame of the top photo)

You can export output signal to the monitor using the HDMI cable from the video camera.

HDMI Cable

BASIC ADJUSTMENT OF MICROSCOPE

Firstly, please check the proper eye position. This eye position is called eye point.

With the old microscope, the eyelashes were close to touch the eyepiece. Now, there are various high eye point that you can observe while wearing glasses.

 

If the eyepiece contains the following marks, it is a high eye point.

 

I will show you one way to easily know eye points.

Place copy paper on the eyepiece.

Changing the distance of the copy paper from the eyepiece changes like the photo below.

The point with the most focus is the eye point.

< Focus status> < Unfocus status>

Place your eyes at the correct position (eye point) then adjust.

1.Inter-eye adjustment (Adjust the eyepiece and eye width.)

When looking at the binocular microscope, there is one tip is to look far away.

Looking into the eyepiece while consciously looking far away then adjust the width of both eyepiece lenses.

 

I looked through the eyepiece with both eyes and the adjustment is complete once the field of view becomes one full circle.

2.Diopter adjustment

Adjust by turning the diopter adjusting ring so that both eyes are in focus.

(One-sided diop adjustment) (Two-sided diop adjustment)

3.Zoom adjustment

(1) Turn the zoom dial to set the minimum magnification and focus.

(2) Next, turn the zoom dial to set the maximum magnification and focus.

(3) Return to low magnification again. Adjustment is completed if the focus is on the low magnification side.

When the focus is out of focus

Repeat steps (1) and (2) for one-sided diop adjustment.

If it is a two-sided diop adjustment, turn the diopter adjusting ring so that the two eyes are in focus at the stage (3).

Then repeat the operations from step (1) to (3).

What’s the difference between optical microscope magnification and digital microscope magnification?

The magnification of the optical microscope and the magnification of digital microscope are very different.

The magnification of optical microscope is absolute magnification, even if the manufacturer differs, if the magnification is the same, it will be in the same field of view. (It is an absolute magnification.)

However, for digital microscope, the field of view will be very different depending on the manufacturer even at the same magnification. (Relative magnification.)

Optical Microscope magnification

The magnification of the microscope is calculated by the following formula.

Optical magnification = (magnification of the objective lens) × (magnification of the eyepiece) Magnification of the microscope = optical magnification.

(In the case of the microscope of the bottom photograph, the magnification becomes 10 times the magnification of the objective lens × the magnification of the eyepiece 10 times, the optical magnification becomes 100 times.)

顕微鏡の倍率

 

Even in the case of a microscope, when the camera is attached to microscope, it may be expressed including the monitor magnification.

(Refer to “How to calculate magnification of camera for microscope”.)

Digital Microscope magnification

Taking an example of a microscope with an objective lens, which is easy to understand.

Firstly, the calculation of optical magnification is the same as the microscope.

Optical magnification = (magnification of the objective lens) × (magnification of the lens)

(In the microscope of the below picture, the magnification of the objective lens is 10 × magnification × 2 times the magnification of the lens body, the optical magnification is “20 times.”)

マイクロスコープの倍率

 

However, the monitor magnification is shown here.

Overall magnification = monitor magnification × optical magnification.

If the image sensor size is 1/2 inch and the monitor is 17 inches, the monitor magnification is 54.

(Please refer to “magnification of microscope”.)

Overall magnification = monitor magnification × optical magnification = 54 × 20 = 1080 times.

In conclusion, if the monitor size or image sensor size changes, the magnification will also change.

(Looking at the catalog of each manufacturer, there is a note such as “Equivalent to” ○○ inches monitor “

< Magnification definition>

Computing is complex, but the way of thinking is simple.

Microscope as shown below with 1mm on the display to 40mm.

This is simply 40 times.

倍率の考え方

 

 

Summary: 

 

The magnification concepts of a microscope and a macro lens are different and not interchangeable.

 

The magnification of a microscope is calculated as: 

Optical Magnification = (Objective Lens Magnification) × (Eyepiece Lens Magnification).

 

The magnification of a macro lens is calculated as:

Optical Magnification = (Objective Lens Magnification) × (Lens Body Magnification).

 

In the case of a microscope, the overall magnification includes the consideration of monitor magnification in addition to the optical magnification.

 

Total Magnification = Monitor Magnification × Optical Magnification.

 

What’s is Gige Camera?

GigE camera is short of Gigabit Ethernet camera.

Expensive dedicated board is not required, general-purpose Ethernet cable can be used for the cable.

The connection port is also LAN port.

However, a power supply for driving the camera is necessary. (See “LAN port connecting GigE camera”)

 

 

 

 

 

The cable can be extended up to 100 m.

Because it is cheap and can communicate at high speed, it is a user-friendly camera that can construct an inexpensive image processing system.

 

The price is equivalent to USB cameras, but the feature of this product is the ability to extend the cable up to 100 meters. Additionally, it offers stable communication.

 

The initial setup (environment configuration) is slightly more complex than USB cameras, but once connected, the communication stability surpasses that of USB.

 

About Image Size of Camera

Cマウントカメラの仕様書にはかならず、「イメージサイズ」、「撮像素子サイズ」が記載されています。
同じレンズを装着した場合、素子サイズが小さい方が視野が狭くなります。
また、このサイズは、視野の計算(レンズの選択)時に必要な値となります。
(視野の計算方法は「レンズのf値」を御参照ください。)

1画素のサイズもこれで計算できます。
例えば、130万画素(1280 X 1024)の1/2.5インチの撮像素子であれば
水平方向  5.6mm/1280画素=0.0044mm/dot
垂直方向  4.2mm/1024画素=0.0041mm/dot   となります。

ちなみに、レンズ側の仕様書にも、「1/2インチカメラ対応」等の記載があります。
この場合、「1/2インチカメラだけに対応している」という意味ではなく、「1/2インチ以下のカメラならば使える」という意味です。
但し、上にも記載したとおり、視野は変わってしまいます。

弊社が主に使っている  1/2インチ、1/2.5インチ、1/3インチ 撮像素子のサイズは下記の通りです。

撮像素子のサイズ

不思議なことに気が付きませんか?
例えば1/2インチ素子の場合、対角が8mmになっています。
1/2インチは12.7mmであるにも関わらず8mmです。
これは撮像素子の特殊な事情で、撮像管時代の基準がそのまま使われているためです。

What is S Mount?

S-Mount với kích thước nhỏ và giá thành thấp.

There is no regulation of flange back.

S-mount are used for compact cameras and board cameras. Currently, there are many lenses with small size and low price.

(例)

 

S → C mount conversion ring is also on sale.

 

C mount has provisions of flange back, whereas S mount does not have flange back specification.

Therefore, even if the aperture is changed, It doesn’t mean that C mount lenses can be used.

When using the C mount lens, I think that it is necessary to try it.

DISTRIBUTION OF METALLURGICAL MICROSCOPES

■Adjust the light source position 金属顕微鏡
(BEFORE ADJUST) (AFTER ADJUST)
金属顕微鏡 金属顕微鏡
   
■ Adjust lens position of the light source 金属顕微鏡
(BEFORE ADJUST) (AFTER ADJUST)
金属顕微鏡 金属顕微鏡
   

■ The image sensor size of camera

Even the image sensor size of camera changes.

Reducing the image sensor size, the influence of variations in illuminance distribution may be reduced. However, its picture will be enlarged than the actual naked eye observation.

Camera sensor 1/2 inch  Camera sensor 1/3 inch
金属顕微鏡 金属顕微鏡
(Note) Since the cameras are different, color and contrast are slightly different.

 

How to see large objects with a microscope?

With normal microscope, the distance from the lens center to microscope stand is limited. Thus, if it is a wide object, it will hit the stand.

 

At our company, we have options (special order) like the theone on the right picture.

顕微鏡
   

This option can be installed both vertically and horizontally on our large stand.

顕微鏡 顕微鏡
   
■Set up on the long side  
顕微鏡

note: When using in this direction we recommend attaching a weight (optional) to stabilize

   
■Set up on the short side  
顕微鏡

 

Proposal for fluorescence (GFP) filter equivalent

Considering the filter that cuts the wavelength of illumination (470 nm, 505 nm) and allows only the emission wavelength to pass. It will be below.
GFPフィルター

SC 56 is Fuji Film’s model number.

The equivalent one is model O56 of HOYA company.

We can install it to our stereo microscope too.

GFP

It is not specified as GFP, but the characteristics are equivalent.

How to attach translucent light to microscope without light?

We will show you how to attach translucent light to the base without transillumination afterwards.

(There are some conditions based on base shape.)

透過照明

As mentioned above, there are the following methods as long as it is a microscope with an observation board (black and white plate).

Firstly, remove the observation board.

透過照明

Place the surface emitting light under the base. Pull out the cable from the gap.

透過照明

Attach the glass plate instead of the observation board.

顕微鏡

Adding a glass plate and surface emission lighting can be used as a simple transparent lighting stand.

 

透過照明   透過照明

 

<Transmitted lighting used>

透過照明   Light illumination for small camera stand RD-95T

 

RMS STANDARD

The RMS (Royal Microscopical Society) standard is the specification of the objective lens connection part (M20.32 mm P = 0.7.6 mm) of the microscope.

 

RMS → C mount

RMS → T mount

 

Conversion adapters such as are commercially available.

You can do various things by using this.

How to take a photograph by using microscope with a scale?

There are scales on the software.

The pitch of the scale can be changed

The above picture is adjusted to divide the 0.1 mm width into 25 equal parts. (You can not put numbers on the scale.)

 

(1)Shooting through a reticle (glass scale) lens with a small digital camera

 

There is also a way to use ordinary digital camera and camera bracket.

 

Using this bracket, and attach a digital camera to the eyepiece of the microscope and shoot.

 

 

 

 

(2) Shooting with a microscope camera that covers the eyepiece

 

Place a eyepiece (glass scale) of the microscope.

Cover the eyepiece with a microscope camera that covers the eyepiece.

 

 

Then you can take pictures with a reticle (glass scale)

Tips for taking pictures with microscopes by single lens reflex camera

Focusing

Focusing during microscope shooting is not a microscope eyepiece but please go while watching the camera’s LCD screen. (It will be a viewfinder on cameras without live view function.)

*auto focus function of camera doesn’t work at this time. so please adjust a focus with using focus angle of stereo microscope.

 

Even if you focus on an eyepiece, it usually does not match the focus of the camera.

(If there is diopter adjustment or back focus on the microscope, tonality may be possible.)

Vibration reduction

It will be like a picture blurred when you shake hands.

To prevent camera vibration, please use remote control or set the time settings.

For cameras with low vibration mode, set the low vibration mode to ON (about 1 second).

Exposure adjustment

The exposure priority mode (dial A) is easy when shooting. If the monitor is too dark or too bright, please adjust the light / dark control (±).

Depending on the camera, “lens unmounted” may become occasionally.

In this case, shoot in manual mode.

(In this case, the brightness is adjusted by the shutter speed.)

White balance

Please manually adjust if the image becomes reddish or blue-colored due to the color of the light source of the microscope.

Make the light source of the microscope proper, remove the specimen, set it to nothing on the screen (glass part without specimen in the case of a biological microscope, white or gray board in case of a stereomicroscope), set the screen to white I will let you recognize.

※ Please see the operation manual of the camera for detailed operation.

ISO sensitivity

If the ISO sensitivity is too high, the image quality will degrade. If the screen is dark, you can take a beautiful image by using some method such as brightening the light of the microscope or slowing down the shutter speed, setting the sensitivity to a lower setting and shooting.

MAGNIFICATION AND FIELD OF VIEW OF CAMERA FOR MICROSCOPE DEFINITIONS

When you observe the microscope with the naked eye, the field of view gets rounded.

(We are using glass scales in increments of 0.5 mm as test samples.)

 

 

 

When this is shot with a camera, it becomes a 4: 3 rectangle with enlarger size.

Besides that, the field of view will change depend on the magnification of the camera connection (connector lens), the size of the image sensor of the camera, and so on. (It will also be a red field of view and a blue field of view in combination.)

You can express this with monitor magnification. (You can also calculate.)

Camera connection part is C-mount

If the camera connection part on the microscope is C mount, the C mount part also has magnification.

If you do not specify microscope, it seems that a C mount adapter of equal magnification (1 times) is often attached.

Monitor magnification = magnification of objective lens × magnification of C mount × (diagonal length of monitor / diagonal length of camera sensor angel)

 

1 / 2.5 inch camera (diagonal length of camera sensor angel is 7 mm)

19 inch monitor (diagonal angel 470 mm),

C mount part (equal magnification),

Objective lens 10 times

<Magnification caculation> 

10 times X 1 times X (470 mm / 7 mm) = 670 times

The field of view will be like the bottom picture.

 

 

Since the horizontal dimension of the 19 inch monitor is 370 mm. So, the image on the right spreads to the screen

<Measured magnification>

370 / 0.55 mm = 672 times

 

Magnification caculation (monitor magnification) and actual magnification are the same.

We need to reduce the magnification of C-mount in order to increase the field of view

We also offer equal size (standard equipment) and 0.5 times (option).

There are various magnifications when looking at other company’s catalogs.

If you are connected to microscope by JIS len, the magnification can not be changed (equal magnification)

If you are connected a camera to this lens barrel, a relay lens is attached to the camera side. There is magnification in this relay lens.

 

Magnification calculation is as follows:

Monitor magnification = magnification of the objective lens × magnification of the connector lens × (diagonal length of monitor / diagonal length of camera sensor angle)

 

1 / 2.5 inch camera (image sensor diagonal 7 mm),

19 inch monitor (diagonal 470 mm)

Connector lens (0.45 times),

Objective lens 10 times

 

<Calculation magnification>

10 times X 0.45 times X (470 mm / 7 mm) = 302 times

Since the horizontal dimension of the 19 inch monitor is 370 mm. So, the picture on the right spreads to the screen then:

<Measured magnification>

370 / 1.2 mm = 308 times

Magnification caculation (based on magnification on monitor) and actual magnification are the same.

Our company has 0.45 times and 0.37 times connector lens available.

We’ll advise the suitable one to the customer according to the size of the image sensor of the camera.

How to save a wide field of view with a camera of microscope

You can see the visual field of the naked eye observation of the microscope is the right picture   顕微鏡の裸眼観察の視野
     

Normally, a 1/2 inch camera and a 0.45-megapixel lens will be included in the kit.

In this case, the field of view will be as shown on the right.

With a connector len of 0.45 times, vignetting (shadow) does not occur at the four corners of the image.

  0.45倍リレーレンズ
     

In case you really want to get a wide field of view, if you attach a wide 0.37x relay lens (wide angle) to the 1/2-inch camera, it will look like the right picture.

* Slightly, ambient distortion occurs.

  0.37倍リレーレンズ
     
    リレーレンズ

How to prevent reflection of lighting by binocular microscope

For monocular len microscope, there are indirect lighting devices such as dome lighting and arch lighting.

間接照明

However, for a binocular stereomicroscope, it is structurally two light path designs.

Therefore, the lighting device like the one above can not be used as it is.

間接照明

There are also a way to weaken the reflection by using the LED (flat type) with the LED pointing straight down as below.

間接照明

For simplicity, there is also a method to observe by attaching a diffusion tube to the LED ring light

間接照明

How to attach general-purpose LED ring light to stereo microscope?

When attaching general-purpose LED ring light (not special LED ring light) to the tip of a stereo microscope, if the inner diameter of LED ring light is larger than the tip part of the microscope (the red circle part of the below picture), basically it can attach.

 

(Microscope with attachment part at the tip)

 

Based on the type ofmicroscope, there is a T-bar hanging at the head or the head is larger than the inner diameter of the LED ring light.

 

(Microscope without attachment part at tip)

 

In such a case, you can easily attach the lighting by attaching the LED fixing ring using the inner screw of the microscope.

For microscope as described above, the inner screw for fixing the filter is cut on the back of the tip cover.

As it is for optical filters, pitch microscope is common regardless of microscope manufacturer. (P = 0.75 mm)

* By the way, it is a fine pitch than fine thread of he general ISO screw.

 

This can be used to attach the fixing ring.

(We have a fixed ring of M49 and M48 available.)

 

     

It is easy to attach LED ring light.

 

 

The inner thread diameter may be large with a very large microscope.

There is no problem in this case.

Step-down rings are on sale from filter makers and others.

 

The following example is a step down ring from M 58 to M 49 mm.

 

 

By combining an LED fixing ring (attach with step down ring M49), you can easily fix the LED ring light to a microscope with a large aperture.

WHAT’S XY TABLE?

A support table (sample table) with a handle that can be moved in X and Y directions. It is advantageous when using the appropriate position adjustment becomes easier.

There are several methos to move the XY table.

1. Direct acting gear (that moves the XY table using the steering wheel)

 

簡易XYテーブル(落射照明タイプ)
簡易XYテーブル(透過照明タイプ)

Simple XY table (Transparent light)

TK100-N

 

2. XY table using micrometers

You can move precisely. It is an advantage to be able to measure the amount of movement.

マイクロメータ付きXYテーブル
デジタルマイクロメータ付きXYテーブル

3.XY table without knob

Since the steering wheel does not protrude, it is possible to place big observed objects.

Sometimes it is called a slide table.

スライドテーブル(特注品) Sliding table
(Special order)

 

※Sliding table movement

LUNAR NEW YEAR HOLIDAY

Shodensha Vietnam is pleased to announce our schedule of Lunar New Year holiday as follow:
-Our office will be closed from Saturday,  10th February 2018 to Tuesday, 20rd Februry 2018
-Business operation will resume as normal on February 21th2018 (Wednesday).
 
We would like to take this opportunity to thank you for your continuous support.

HAPPY NEW YEAR HOLIDAY

Thank you for your trust, using our products and cooperate with us for a long time.
Shodensha Vietnam is pleased to announce our schedule of Happy New Year holiday as follow:
 
-Our office will be closed from Saturday,  30 th December 2017 to Wednesday, 3rd January 2018
-Business operation will resume as normal on Thursday 4th2018 (Tuesday).
 
Wishing you and your family a happy holiday.

DEFINITION OF MICROSCOPE MAGNIFICATION

The definition of microscope magnification is quite simple.
For eg: If the observed object (1mm) is displayed on the PC monitor or PC to 10mm, the magnification will be 10 times

Caculation:

General magnification = monitor magnification x optical magnification (lens scale)

Monitor magnification = monitor size (inch) x 25.4mm/flange distance

Flange distance 6mm (1/3” camera), 7mm (1/2.5” camera), 8mm (1/2” camera) (Please refer to [Image size of camera] for more information)

The magnification of the lens was calculated value when it was attached to our 1/2 inch camera (variable magnification lens for coaxial lighting SDS-FZ)  and observed with a 17-inch monitor.

Monitor magnification = (17×25.4)/8 = 54

General magnification = 54×4 = 216 times

高解像度高倍率ズームレンズ SDS-FZR
   

By combining camera to auxiliary lens, ring lens, conveter lens, objective lens, …, the magnification will be increased.

The magnification of the lens was calculated value when it was attached to our 1/2 inch camera (Standard variable magnification lens SDS-M) and observed with a 17-inch monitor.

Monitor magnification = (17×25.4)/8 = 54

General magnification = 54x4x0.5 (lens magnification) = 108 times

ズームレンズ SDS-M
   

 

■ How to show filed of view of storaged images?

The magnification of the microscope has a lot of parameters, it becomes complicated when writing reports.

Therefore, please keep the calibration value instead of magnification that shown in the right picture.

Then you can determine the approximate field of view of the image.

計測ソフトMF Ship
   

Exhibition Announcement

Shodensha Vietnam is going to participate in an exhibition, Metalex Vietnam 2017,

held at Saigon Exhibition and Convention Center(SECC)on Oct. 12th – 14th.

 

 

We will hand over a 15% discount card※ and a small loupe in exchange for a business card.

※Except for some items. There is an expiration date.

 

◆◇◆INFO of Metalex Vietnam 2017◆◇◆

Name: Metalex Vietnam 2017 (Vietnam’s international exhibition on machining tools and metalworking solution for production upgrade)

Date: Oct. 12th – 14th 2017

Hours: 9AM – 5PM Daily

Place: Saigon Exhibition and Convention Center(SECC)

Booth: B 08 (Hall A)

NATIONAL HOLIDAY ANNOUNCEMENT

Thank you for your trust, using our products and cooperate with us for a long time.

Shodensha Vietnam is pleased to announce our schedule of Independence Day holiday 02nd September 2017 as follow:

 

-Our office will be closed from Saturday,  2ndSeptember 2017 to Monday, 4th September 2017

-Business operation will resume as normal on September 5th2017 (Tuesday).

 

Wishing you and your family a happy holiday.